ALCOA+Data-integrity principles: Attributable, Legible, Contemporaneous, Original, Accurate, plus Complete, Consistent, Enduring and Available.
Analytical Target Profile (ATP)A prospective statement (ICH Q14) of what the method must measure, over what range, and to what performance — the specification a method is developed and validated against.
BackpressureThe pressure the pump must generate to push the mobile phase through the system. Proportional to viscosity, column length and flow, and inversely proportional to the square of particle size.
Band broadeningThe spreading of an analyte band as it travels. Variances from the column, injector, tubing and detector add: σ²_total = Σσ².
Capacity factorOlder name for the retention factor k.
CarryoverAnalyte from one injection appearing in the next. Typically limited to < 0.1% of the main peak for a chromatographic assay, or ≤ 20% of the LLOQ response in bioanalysis.
Core-shell particleA solid core with a porous outer layer. Shortens the diffusion path, giving near-sub-2-µm efficiency at conventional pressure.
Critical pairThe adjacent peak pair with the smallest resolution — the pair that determines whether the method works.
Dwell volumeVolume between the gradient mixing point and the column head. Delays the gradient's arrival and is the leading cause of failed gradient method transfer.
Efficiency (N)Theoretical plate count, N = 16(t_R/W)² or 5.54(t_R/W½)². A measure of how narrow a peak is relative to its retention time.
EndcappingReaction of residual surface silanols with a small silane after the main bonding step, to reduce tailing of basic compounds.
Extra-column volumeAll the volume outside the column packing — needle seat, tubing, fittings, flow cell. Critical on narrow-bore columns.
Forced degradationDeliberate stressing (acid, base, oxidation, heat, light, humidity) to generate degradants and demonstrate that the method is stability-indicating. Target 5–20% degradation.
GradientA programmed change in mobile-phase composition during the run, used when the analytes span too wide a polarity range for one isocratic strength.
Guard columnA short cartridge upstream of the analytical column that traps particulates and strongly retained matrix.
HILICHydrophilic interaction liquid chromatography — a polar stationary phase with a high-organic mobile phase, retaining polar analytes that reversed phase cannot hold.
IsocraticConstant mobile-phase composition throughout the run.
Linear solvent strength (LSS)log k = log k_w − S·φ. The relationship underlying nearly all reversed-phase retention prediction.
LOD / LOQDetection limit (≈ S/N 3:1) and quantitation limit (≈ S/N 10:1). The LOQ must be confirmed experimentally with acceptable accuracy and precision.
Mass balanceIn a stressed sample: assay plus total degradants, typically expected to account for about 95–105% of the initial material.
Mobile phaseThe liquid that carries the sample through the column.
Peak capacity (n_c)The number of peaks that could be baseline-resolved across a gradient; n_c ≈ 1 + t_G/w.
Peak purityA DAD-based check for spectral inhomogeneity across a peak. Evidence, not proof, of a single component.
Plate height (H)H = L/N. Modelled by the van Deemter equation H = A + B/u + C·u.
Reduced plate height (h)h = H/d_p. A well-packed column achieves h ≈ 2–3, independent of particle size — the fair way to compare columns.
Resolution (Rs)Rs = 2Δt_R/(W₁+W₂). 1.5 is baseline separation of equal peaks; ≥ 2.0 is the usual pharmacopoeial default for an assay.
Retention factor (k)k = (t_R − t₀)/t₀. Flow- and dimension-independent; target 2–10.
Reversed phaseNon-polar stationary phase with a polar mobile phase — the mode used for roughly 90% of pharmaceutical HPLC.
RobustnessThe method's tolerance to deliberate small variations in its own parameters. Distinct from ruggedness between laboratories.
Selectivity (α)α = k₂/k₁ for an adjacent pair. The most powerful resolution lever and the only one that can fix a co-elution.
SilanolAn Si–OH group on the silica surface. Weakly acidic; interacts with protonated bases and causes tailing.
SpecificityEvidence that the measured response comes from the analyte alone, with no contribution from matrix, impurities or degradants.
System suitability (SST)A check, run with every sequence, that the whole system is performing adequately today. Without a passing SST the sequence's data has no standing.
Tailing factor (T)T = W₀.₀₅/(2f) at 5% of peak height. USP acceptable range typically ≤ 2.0; Ph. Eur. and JP use the symmetry factor, usually 0.8–1.5.
USP L-codeThe USP classification of a stationary phase (L1 = C18, L7 = C8, L11 = phenyl, L43 = PFP...). A family, not a specification.
van Deemter equationH = A + B/u + C·u. Predicts an optimum linear velocity; sub-2-µm and core-shell particles flatten the high-velocity branch.
Void time (t₀)Elution time of an unretained compound; t₀ = V_M/F with V_M ≈ 0.68·π·(d_c/2)²·L.