37 terms, each defined the way it is actually used at the bench.
| Term | Definition |
|---|---|
| ALCOA+ | Data-integrity principles: Attributable, Legible, Contemporaneous, Original, Accurate, plus Complete, Consistent, Enduring and Available. |
| Analytical Target Profile (ATP) | A prospective statement (ICH Q14) of what the method must measure, over what range, and to what performance — the specification a method is developed and validated against. |
| Backpressure | The pressure the pump must generate to push the mobile phase through the system. Proportional to viscosity, column length and flow, and inversely proportional to the square of particle size. |
| Band broadening | The spreading of an analyte band as it travels. Variances from the column, injector, tubing and detector add: σ²_total = Σσ². |
| Capacity factor | Older name for the retention factor k. |
| Carryover | Analyte from one injection appearing in the next. Typically limited to < 0.1% of the main peak for a chromatographic assay, or ≤ 20% of the LLOQ response in bioanalysis. |
| Core-shell particle | A solid core with a porous outer layer. Shortens the diffusion path, giving near-sub-2-µm efficiency at conventional pressure. |
| Critical pair | The adjacent peak pair with the smallest resolution — the pair that determines whether the method works. |
| Dwell volume | Volume between the gradient mixing point and the column head. Delays the gradient's arrival and is the leading cause of failed gradient method transfer. |
| Efficiency (N) | Theoretical plate count, N = 16(t_R/W)² or 5.54(t_R/W½)². A measure of how narrow a peak is relative to its retention time. |
| Endcapping | Reaction of residual surface silanols with a small silane after the main bonding step, to reduce tailing of basic compounds. |
| Extra-column volume | All the volume outside the column packing — needle seat, tubing, fittings, flow cell. Critical on narrow-bore columns. |
| Forced degradation | Deliberate stressing (acid, base, oxidation, heat, light, humidity) to generate degradants and demonstrate that the method is stability-indicating. Target 5–20% degradation. |
| Gradient | A programmed change in mobile-phase composition during the run, used when the analytes span too wide a polarity range for one isocratic strength. |
| Guard column | A short cartridge upstream of the analytical column that traps particulates and strongly retained matrix. |
| HILIC | Hydrophilic interaction liquid chromatography — a polar stationary phase with a high-organic mobile phase, retaining polar analytes that reversed phase cannot hold. |
| Isocratic | Constant mobile-phase composition throughout the run. |
| Linear solvent strength (LSS) | log k = log k_w − S·φ. The relationship underlying nearly all reversed-phase retention prediction. |
| LOD / LOQ | Detection limit (≈ S/N 3:1) and quantitation limit (≈ S/N 10:1). The LOQ must be confirmed experimentally with acceptable accuracy and precision. |
| Mass balance | In a stressed sample: assay plus total degradants, typically expected to account for about 95–105% of the initial material. |
| Mobile phase | The liquid that carries the sample through the column. |
| Peak capacity (n_c) | The number of peaks that could be baseline-resolved across a gradient; n_c ≈ 1 + t_G/w. |
| Peak purity | A DAD-based check for spectral inhomogeneity across a peak. Evidence, not proof, of a single component. |
| Plate height (H) | H = L/N. Modelled by the van Deemter equation H = A + B/u + C·u. |
| Reduced plate height (h) | h = H/d_p. A well-packed column achieves h ≈ 2–3, independent of particle size — the fair way to compare columns. |
| Resolution (Rs) | Rs = 2Δt_R/(W₁+W₂). 1.5 is baseline separation of equal peaks; ≥ 2.0 is the usual pharmacopoeial default for an assay. |
| Retention factor (k) | k = (t_R − t₀)/t₀. Flow- and dimension-independent; target 2–10. |
| Reversed phase | Non-polar stationary phase with a polar mobile phase — the mode used for roughly 90% of pharmaceutical HPLC. |
| Robustness | The method's tolerance to deliberate small variations in its own parameters. Distinct from ruggedness between laboratories. |
| Selectivity (α) | α = k₂/k₁ for an adjacent pair. The most powerful resolution lever and the only one that can fix a co-elution. |
| Silanol | An Si–OH group on the silica surface. Weakly acidic; interacts with protonated bases and causes tailing. |
| Specificity | Evidence that the measured response comes from the analyte alone, with no contribution from matrix, impurities or degradants. |
| System suitability (SST) | A check, run with every sequence, that the whole system is performing adequately today. Without a passing SST the sequence's data has no standing. |
| Tailing factor (T) | T = W₀.₀₅/(2f) at 5% of peak height. USP acceptable range typically ≤ 2.0; Ph. Eur. and JP use the symmetry factor, usually 0.8–1.5. |
| USP L-code | The USP classification of a stationary phase (L1 = C18, L7 = C8, L11 = phenyl, L43 = PFP...). A family, not a specification. |
| van Deemter equation | H = A + B/u + C·u. Predicts an optimum linear velocity; sub-2-µm and core-shell particles flatten the high-velocity branch. |
| Void time (t₀) | Elution time of an unretained compound; t₀ = V_M/F with V_M ≈ 0.68·π·(d_c/2)²·L. |
Produced by Mirela Gorjanu · Practical Pharma Science™.