Detection: what the peak really is
The detector converts a concentration profile into a signal. Its wavelength, its bandwidth, its data rate and its linear range all decide whether your number is real.
A UV detector obeys Beer–Lambert: A = ε·b·c. Response is proportional to concentration, and the constant of proportionality (ε) is wavelength- and compound-specific — which is why you cannot compare peak areas of different compounds without a response factor.
Choosing the wavelength
- For an assay: use λmax for maximum sensitivity and the flattest response to small wavelength errors.
- For related substances: choose a wavelength where the impurities and the main compound have similar absorptivity, even if it is not anyone's λmax — otherwise a 0.1% area is not a 0.1% mass.
- Avoid the steep flank of an absorbance band: a ±2 nm detector error there changes response by several percent.
- Below 220 nm, every mobile-phase component absorbs. Check the UV cut-off of your buffer and modifier before blaming the sample.
| Setting | Symptom if wrong | Rule of thumb |
|---|---|---|
| Data rate too low | Peaks look short and broad; area low | ≥ 20 points across the narrowest peak |
| Response time / filter too long | Peaks distorted, resolution lost | Match to peak width; shorten for UHPLC |
| Reference wavelength badly chosen | Baseline drift, negative dips | Use a region with no analyte absorbance, or turn it off |
| Flow-cell volume too large | Extra-column broadening on narrow-bore columns | ≤ 10% of the peak volume |
| Absorbance too high | Non-linear calibration at the top standard | Keep peak apex below ~1.0–1.5 AU |
Linear range is a property of the detector AND the concentration, not of the compound. If your highest standard is above about 1.2 AU you will see a flattening calibration curve and a failed linearity criterion. Dilute, do not fit a quadratic to hide it.
Other detectors you will meet
| Detector | Detects | Strength | Limitation |
|---|---|---|---|
| UV / DAD | Chromophores | Robust, linear, cheap, spectra for peak purity | Blind to non-chromophoric analytes |
| Fluorescence | Fluorophores | 10–1000× more sensitive and very selective | Few analytes; quenching effects |
| RI | Any refractive-index difference | Universal | Insensitive; isocratic only; temperature-critical |
| ELSD / CAD | Non-volatile analytes | Universal, works without a chromophore | Non-linear response; needs volatile buffers |
| MS | Mass-to-charge | Identity plus sensitivity plus selectivity | Volatile buffers only; ion suppression; cost |
Peak purity from a DAD is evidence, not proof. It can only detect a co-eluting impurity whose spectrum differs and whose concentration is high enough to shift the observed spectrum. A clean purity result never replaces an orthogonal method or an MS check.
Your assay is linear to 80% of the working range and then bends. The top standard's peak reaches 1.8 AU. What is happening?
HPLC Course Pro™ is an educational training tool produced by Mirela Gorjanu · Practical Pharma Science™. The simulators are teaching models, not predictive engines: every condition must be confirmed experimentally. This application is not a validated computerised system, does not generate a controlled GxP record, and is not a system of record. Pharmacopoeial criteria quoted here are general-chapter defaults; the applicable monograph or your own validated method always takes precedence.